Accuracy assessment of the blood glucose meter «Satellite Express» for glycemic control in children with diabetes
https://doi.org/10.14341/probl12311
Abstract
Background: Monitoring glycemic control in diabetes includes daily measurements of glucose levels at home, along with periodic monitoring of glycemia indicators on the whole. The main purpose of monitoring glycemic control is to accurately and authentically assess the glycemic level reached by each patient in order to ensure the glycemic goals achievement. Glycemic control assessment at a certain point in time is best achieved by self-monitoring the level of glucose in the blood, in that allows you to assess the presence of hyperglycemia and hypoglycemia to optimize the treatment strategy.
Aims: Accuracy assessment of a medical device «Portable blood glucose concentration meter with electrochemical single-use strip PKG-03 «Satellite Express» for glycemic control of children with diabetes.
Materials and methods: In a group of children with diabetes (n=105, age from 1 month to 17 years), simultaneously was taken capillary blood from one drop to assess the accuracy of glycemia levels on the «Satellite Express» glucose meter and on the laboratory glucose and lactate analyzer SUPER GL.
Results: In a comparative analysis of 660 blood samples on the Clark error scale, was found that all deviations of glycemia values obtained with the «Satellite Express» glucose meter were in zones A (99.7% — clinically correct values) and B (0.3% — safe deviations). The limits of permissible systematic error of measurements that meet the requirements of the National Standard of Russian Federation GOST R ISO 15197-2015 in terms of clause 6.3 System Accuracy amounted to 96.8% of deviations of glucose meter indications from reference values, which matches to areas of clinically correct and safe deviations.
Conclusion: Using the «Satellite Express» glucose meter in clinical practice makes it possible to control glycemia qualitatively and it is safe for children aged from 1 month to 17 years.
About the Authors
Oleg A. DianovRussian Federation
MD, PhD
Vadim V. Maltsev
Elena S. Murina
Ilya S. Anfilov
References
1. Jansen RT, Slingerland RJ. SKML-Quality Mark for point-of-care test (POCT) glucose meters and glucose meters for home-use. Clin Chem Lab Med. 2010;48(7):1021-1027. doi: 10.1515/CCLM.2010.226.
2. Ricos C, Alvarez V, Cava F, et al. Current databases on biological variation: pros, cons and progress. Scand J Clin Lab Invest. 1999; 59(7):491-500. doi: 10.1080/00365519950185229.
3. ГОСТ Р ИСО 22870-2009. Исследования по месту лечения – Требования к качеству и компетентности. — М.: Стандартинформ; 2010. [GOST R ISO 22870-2009. Research at the treatment site – Quality and competency requirements. Moscow: Standartinform; 2010. (in Russ).]
4. Freckmann G, Schmid C, Baumstark A, et al. System Accuracy Evaluation of 43 Blood Glucose Monitoring Systems for Self-Monitoring of Blood Glucose according to DIN EN ISO 15197. J Diabetes Sci Technol. 2012;6(5):1060-1075. doi: 10.1177/193229681200600510.
5. Дедов И.И., Шестакова М.В., Майоров А.Ю., и др. «Алгоритмы специализированной медицинской помощи больным сахарным диабетом» Под редакцией И.И. Дедова, М.В. Шестаковой, А.Ю. Майорова 9-й выпуск // Сахарный диабет. – 2019. – Т. 22. – №1S1. – C. 1-144. [Dedov II, Shestakova MV, Mayorov AY, et al. Standards of specialized diabetes care. Edited by Dedov I.I., Shestakova M.V., Mayorov A.Yu. 9th edition. Diabetes mellitus. 2019;22(1S1):1-121. (in Russ).] doi: 10.14341/dm221s1
6. Пашинцева Л.П., Бардина В.С., Ильясов И.Р. и др. Клинико-лабораторная оценка точности портативных глюкометров «Сателлит экспресс» и «Сателлит экспресс мини» // Клиническая лабораторная диагностика. — 2011. — № 11. — С. 33-35. [Pashintseva LP, Bardina VS, Ilyasov IR, Mischenko BP, Antziferov MB. The clinical laboratory evaluation of accuracy of portable glucometers Satellite Express and Satellite Expressmini. Russian Clinical Laboratory Diagnostics. 2011;(11):33-35. (in Russ).]
7. ГОСТ Р ИСО 15197-2015. Тест-системы для диагностики in vitro – Требования к системам мониторинга глюкозы в крови для самоконтроля при лечении сахарного диабета. — М.: Стандартинформ; 2015. [GOST R ISO 15197-2015. In vitro diagnostic test systems – Requirements for blood glucose monitoring systems for self-monitoring in the treatment of diabetes mellitus. Moscow: Standartinform; 2015. (in Russ).]
8. Курникова И.А., Уалиханова А.У., Моргунов Л.Ю., и др. Оценка эффективности использования средств гликемического контроля в условиях неудовлетворительной компенсации сахарного диабета. // Проблемы эндокринологии. — 2017. — Т. 63. — №1. — С. 23-29. [Kurnikova IA, Ualihanova AU, Morgunov LY, Mavlyalieva ER, Surikova MA. Assessment of efficiency of glycemic control agents in insufficient compensation of diabetes. Problems of endocrinology. 2017;63(1):23-29.] doi: 10.14341/probl201763123-29
9. Self-monitoring of blood glucose. American Diabetes Association. Diabetes Care. 1994;17(1):81-86. doi: 10.2337/diacare.17.1.81
Supplementary files
|
1. Рисунок 1. Оценка клинической точности глюкометра «Сателлит Экспресс» по Согласованной сетке ошибок показателей капиллярной крови. | |
Subject | ||
Type | Other | |
View
(2MB)
|
Indexing metadata ▾ |
|
2. Рисунок 2. Зависимость показателей капиллярной гликемии, измеренной с помощью глюкометра «Сателлит Экспресс» и референтным методом | |
Subject | ||
Type | Other | |
View
(1MB)
|
Indexing metadata ▾ |
|
3. Рисунок 3. График точности системы «Сателлит Экспресс» | |
Subject | ||
Type | Other | |
View
(1MB)
|
Indexing metadata ▾ |
|
4. Fig. 1. Assessment of the clinical accuracy of the Satellite Express glucometer according to the Agreed grid of capillary blood count errors. | |
Subject | ||
Type | Other | |
View
(1019KB)
|
Indexing metadata ▾ |
|
5. Fig. 2. The dependence of capillary glycemia indicators, measured using a satellite express glucometer and the reference method. | |
Subject | ||
Type | Other | |
View
(918KB)
|
Indexing metadata ▾ |
|
6. Fig. 3. The accuracy chart of the Satellite Express system. | |
Subject | ||
Type | Other | |
View
(680KB)
|
Indexing metadata ▾ |
Review
For citations:
Dianov O.A., Maltsev V.V., Murina E.S., Anfilov I.S. Accuracy assessment of the blood glucose meter «Satellite Express» for glycemic control in children with diabetes. Problems of Endocrinology. 2019;65(6):436-443. https://doi.org/10.14341/probl12311