Evaluation of the efficiency of using glucose monitoring devices upon unsatisfactory diabetes compensation
https://doi.org/10.14341/probl201763123-29
Abstract
Rationale. The key purpose of modern glucose meters is to ensure regular self-monitoring of glucose level when receiving outpatient management, provided that fair diabetes compensation is achieved. Although glucose meters are not intended for assessing the glucose level in severe metabolic disorders (ketosis, ketoacidosis), since these conditions have a negative effect on device accuracy, in actual life a patient (or a physician) can face a situation when a glucose meter is the only tool for evaluating carbohydrate metabolism disorders.
Objective. To evaluate clinical accuracy of Satellite Express PKG-03 glucose meter in measuring the glucose level in patients with type 1 and 2 diabetes mellitus (DM) receiving insulin therapy when the disease course is complicated by ketosis or ketoacidosis.
Material and methods. Capillary blood was simultaneously collected in two groups of patients receiving insulin therapy from the same drop to evaluate the glucose blood level using the Satellite Express glucose meter and a SUPER GL laboratory analyzer of glucose and lactate levels. Acid-base imbalance was the key criterion for distributing patients into groups: no disorders were detected in group 1 patients, while group 2 patients had ketosis or ketoacidosis. The results were evaluated using the Clarke error grid.
Results. Comparative analysis of blood samples collected from 77 patients showed that all deviations in glucose level indices measured using the Satellite Express glucose meter from the reference values belonged to zones A (the clinically valid values) and B (safe deviations) in patients without acid-base imbalance. In patients hospitalized for ketosis and ketoacidosis (group 2), the deviations from the reference values lay in zones A and B in 97%, while lying on the boundary between zones B and C only in 3%.
Keywords
About the Authors
Irina A. KurnikovaPeople’s Friendship University of Russia
Russian Federation
M.D. DMS Faculty, Department of hospital therapy with courses of endocrinology, a hematology and clinical laboratory diagnostics
Competing Interests:
no
Aigerim U. Ualihanova
People’s Friendship University of Russia
Russian Federation
graduate student Department of hospital therapy with courses of endocrinology, a hematology and clinical laboratory diagnostics
Competing Interests:
нет
Leonid Y. Morgunov
People’s Friendship University of Russia; A.K. Eramishantsev Municipal Clinical Hospital
Russian Federation
M.D., manager of office of endocrinology
Competing Interests:
нет
Elvira R. Mavlyalieva
A.K. Eramishantsev Municipal Clinical Hospital
Russian Federation
manager of office
Competing Interests:
no
Marya A. Surikova
A.K. Eramishantsev Municipal Clinical Hospital
Russian Federation
manager of laboratory
Competing Interests:
no
References
1. Майоров А.Ю., Мельникова О.Г., Филиппов Ю.И. Самоконтроль гликемии — основа эффективного лечения сахарного диабета. // Эффективная фармакотерапия. эндокринология. — 2014. — Т.46. — №5. — С. 62-72. [Mayorov AY, Melnikova OG, Filippov YuI. Self-monitoring of blood glucose: a mainstay of effective diabetes treatment. Effectivnaya farmakoterapiya. 2014;46(5):62-72. (In Russ).].
2. Дедов И.И., Шестакова М.В., Галстян Г.Р., и др. Алгоритмы специализированной медицинской помощи больным сахарным диабетом. Под ред. Дедова И.И., Шестаковой М.В.(7-й выпуск) // Сахарный диабет. — 2015. — Т. 18. — № 1S. — С. 1-112. [Dedov II, Shestakova MV, Galstyan GR, et al. Standards of specialized diabetes care. Edited by Dedov II, Shestakova MV. (7th edition). Diabetes mellitus. 2015;18(1S):1-112. (in Russ.)]. doi: 10.14341/dm20151s1-112
3. Национальный стандарт Российской Федерации ГОСТ Р ИСО 14155-2014. Клинические исследования. Надлежащая клиническая практика. Введty 2015-06-01 / Федеральное агентство по техническому регулированию и метрологии. МЕДИТЕСТ. — М.: Стандартинформ; 2015. — 47 с. Доступно по http://www.internet-law.ru/gosts/gost/57438/. Ссылка активна на 12.12.2016. [National standard of Russian Federation GOST R ISO 14155-2014. Clinical studies. Good clinical practice. Accepted 2015-06-01 / Federal agency on technical regulating and metrology. MEDITEST. m.: Standartinform; 2015. 47 p. Available on URL: http://www.internet-law.ru/gosts/gost/57438/. (in Russ.)].
4. Freckmann G, Baumstark A, Jendrike N, et al. System Accuracy Evaluation of 27 Blood Glucose Monitoring Systems According to DIN EN ISO 15197. Diabetes Technol Ther. 2010;12(3):221-231. doi: 10.1089/dia.2009.0128 doi: http://dx.doi.org/10.14341/probl201763123-29
Supplementary files
|
1. Рис. 1. Этапы развития технологий гликемического контроля. | |
Subject | ||
Type | Исследовательские инструменты | |
View
(83KB)
|
Indexing metadata ▾ |
|
2. Рис. 2. Оценка клинической точности ПКГ-03 Сателлит Экспресс по согласованной сетке ошибок (Consensus Error Grid, CEG) по показателям капиллярной крови. Здесь и на рис. 3: | |
Subject | ||
Type | Исследовательские инструменты | |
View
(154KB)
|
Indexing metadata ▾ |
|
3. Рис. 3. Оценка клинической точности ПКГ-03 Сателлит Экспресс по согласованной сетке ошибок (Consensus Error Grid, CEG) по анализам нативной крови пациентов, находящихся в состоянии кетоацидоза. | |
Subject | ||
Type | Исследовательские инструменты | |
View
(92KB)
|
Indexing metadata ▾ |
Review
For citations:
Kurnikova I.A., Ualihanova A.U., Morgunov L.Y., Mavlyalieva E.R., Surikova M.A. Evaluation of the efficiency of using glucose monitoring devices upon unsatisfactory diabetes compensation. Problems of Endocrinology. 2017;63(1):23-29. https://doi.org/10.14341/probl201763123-29

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License (CC BY-NC-ND 4.0).